CB-MF-SOP-002
A/1
37
4. Since pipes, wires, etc. need to be connected behind the equipment, cleaning fluid
barrel and liquid waste barrel need to be installed, so the equipment should be
placed to facilitate personnel's access.
5. The instrument will adjust the power supply voltage automatically within the working
voltage range of 100-240 volts. Do not exceed this power supply range.
6. Non-professional personnel shall not open the rear panel of the instrument.
7. The counting chamber is disposable and cannot be reused.
8. Do not touch the unused counter chamber, or any stain may affect the imaging
quality of the equipment.
9. Please do not tear the sealing strip of counting chamber before loading the rotary
disk to prevent the counting chamber plate from falling off.
10. When adding the test tube rack, it must be carried out after the red indicator light on
the feeding platform is off, otherwise the hook pushing the test tube rack will be
overturned when it is returned.
11. When the red indicator light on the feeding module of the test tube is on, it is
forbidden to load the test tube rack on the feeding platform, otherwise, the
instrument will fail or possibly damage it. It is not allowed to be put into the test tube
rack until the red light is off.
12. The used counting chambers should be handled with care, as the urine is derived
from human body fluids and should be treated in accordance with the uniform rules
for hospital biological hazards.
13. Carefully dispose of used counting chambers and urine containers.
14. Wear rubber gloves or other protective clothing when operating S80.
15. Do not spray cleaning inside the equipment. Remove the relevant removable parts
before using a wet cloth to clean the internal parts.
16. Do not move and loosen the counting chamber mounting screw above the
microscope.
17. The fuses on the circuit board of the equipment shall not be replaced unless by
professional maintenance personnel of COBIO Smart Healthcare Technology Co.,
Ltd.
18. Don't touch the microscope lens with your hands. Only use cotton cloth when
cleaning.
19. Do not use any cleaning liquid other than alcohol to wipe the lens. Alcohol should be
anhydrous alcohol and dip by medical cotton balls to be used. Professional
maintenance personnel of non COBIO Smart Healthcare Technology Co., Ltd. are not
allowed to carry out this operation.
20. The equipment must use the counter chamber series of COBIO Smart Healthcare
Technology Co., Ltd., and must be licensed by COBIO.
21. Again carrying and packing the process should pay attention to handle gently,
because the host as the heavy equipment, at least two people carrying, to pay
attention to the safety of personnel and equipment.
22. Only trained personnel can use the equipment.