System Settings
4-10
are available; under auto and rhythm mode, only two kinds of speed [25mm/s]
and [50mm/s] are available.
[Wave Gain]: there are seven wave gains including [AGC], [2.5mm/mV], [5
mm/mV], [10mm/mV], [20mm/mV], [10/5 mm/mV] and [20/10 mm/mV]. In
[20/10 mm/mV], 20 represents the sensitivity of limb leads, and 10 represents
the sensitivity of chest leads, [AGC] means auto gain control.
[Report Format]: when [Printer Select] is set as [External] or [Close], the
available report formats include [3×4], [3×4+1R], [3×4+3R], [6×2],
[6×2+1R], [12×1] and [12×1+1T]; when [Printer Select] is set as [Internal],
available report formats include [3×4] and [3×4+1R].
[3×4]: 12 leads are recorded in 3 channels and 4 sequences, when
“Printer Select” is set as “External” or “Close”, record 2.5s for each
sequence, when “Printer Select” is set as “Internal”, the recoding length
of each sequence is decided by printing length.
[3×4+1R]: 12 leads are recorded in 3 channels and 4 sequences, when
“Printer Select” is set as “External” or “Close”, record 2.5s for each
sequence, when “Printer Select” is set as “Internal”, the recoding length
of each sequence is decided by printing length. And add 1 channel of
rhythm lead waveform.
[3×4+3R]: 12 leads are recorded in 3 channels and 4 sequences, record
2.5 seconds for each sequence and add 3 channels of rhythm lead
waveform.
[6×2]: 12 leads are recorded in 6 channels and 2 sequences, and record 5
seconds for each sequence.
[6×2+1R]: 12 leads are recorded in 6 channels and 2 sequences, record 5
seconds for each sequence and add 1 channel of rhythm lead waveform.
[12×1]: 12 leads are recorded in 12 channels and record 10 seconds all
together.
[12×1+1T]: 12 leads are recorded in 12 channels, in which there is 1
channel of rhythm lead waveform identified with “Rhy.”