Safety Guidance
1-2
1.2.1 Precautions for Operating the Equipment
Warning
This equipment is not intended for treatment.
The use of this equipment is restricted to one patient at a time.
This equipment can not be used directly for the cardiac surgery.
Do not use this equipment in the presence of flammable anesthetics to
avoid explosion.
Do not use this equipment in the presence of high-voltage equipment or
high electrostatic quantity.
Do not posit the equipment to make it difficult to operate the power
plug which uses to isolate the equipment circuits electrically from the
supply mains.
The housing of the equipment should be grounded well. Use the
three-wirereceptacle with protective grounding and the grounding of
the receptacle should be kept well.
This equipment should be installed by the qualified maintenance
engineer. Only the authorized maintenance engineer can open the
housing of the equipment.
If there is doubt for the integrality for the protective ground lead,
please use the built-in battery for the power supply.
The accessory equipments connected with the analog and digital
interfaces of H3 Electrocardiograph should be verified according to
IEC standards. All the configurations should comply with the valid
version of IEC 60601-1. Therefore, the personnel who connect the
accessory equipments to signal input/output interfaces should confirm
the whole system comply with the requirements in the valid version of
IEC 60601-1. If there is any doubt, please contact Comen Company or
local agents.
In order to ensure the patient safety, the summation of leakage current
should never exceed leakage current limits when several other