System Settings
4-12
Select] is set as [External] or [Close], available average template includes
[3×4+1R], [6×2+1R] and [Off].
[3×4+1R]: 12 leads of average template waveform are recorded in 3 channels
and 4 sequences and add one average template waveform of the rhythm lead.
[2×6+1R]: 12 leads of average template waveform are recorded in 2 channels
and 6 sequences and add one average template waveform of the rhythm lead.
[6×2+1R]: 12 leads of average template waveform are recorded in 6
channels and 2 sequences and add one average template waveform of the
rhythm lead;
[Off]: there is no average template waveform output.
[Save Option]: [On] and [Off];
[On]: under auto mode, all the recorded ECG data will be saved in the “File
List” screen automatically.
[Off]: all the recorded ECG data will not be saved.
When the storage is full, the prompt message [MEM FULL] will appear
in the promt message area to remind the user.
[Paper Style]: when [Printer Select] is set as [Internal] or [Close], available
paper style includes [Rolled] and [Folded]; when [Printer Select] is set as
[External], available paper style includes [(A4) Folded 295×210mm], [(16K)
Folded 195×271mm]and [(B5) Folded 182×257mm].
This recorder adopts 80mm×20mm rolled thermosensitive record paper or
80mm×70mmfolded thermosensitive record paper. When no record paper is
loaded or it is used up, [Paper?] prompt message will be displayed to remind the
user of loading or replacing the record paper.