System Settings
4-14
[BMP],[JPG],[PDF],[SCP],[XML] or [DICOM]:The corresponding files of
this format can be saved only when [Printer Select] is set to [Close], and
[Save Option] is on.
[Print function selection]: [One key print], [Extend the record] and [Normal
print];
[One key print]: under auto mode, press key to print the ECG
waveform and report, and after this a rhythm waveform report will be
printed;
[Extend the record]: under auto mode, if any of the following emergencies,
including extreme tachycardia, extreme bradycardia, tachycardia and
bradycardia, occurs in the diagnosis result after measurement finishes, a
rhythm recording will be added automatically.
[Normal print]: under auto mode, only output an ECG report.
[Rhythm R-R Analysis]: click it to activate or deactivate the function of RR
interval analysis.
[Printer Test]: click it to start or stop printer test.
The printer test can detect whether the print head works normally. After loading
record paper, click “Printer Test” option to start printer test, and at this time the
recorder starts to output record paper and print triangle wave in valid 72mm
printing range. The integrality and clarity can be used to judge whether the
thermosensitive print head works normally.
When the recorder is abnormal, it must be maintained or modified by
the qualified engineer designated by Comen Company, otherwise
Comen company shall not be responsible for the safety, reliability and
performance of the instrument.