18
Technical specifications table 24
U.M.
S
B BHD
Steam generation in chamber in chamber in chamber
Chamber capacity l 23,15
Clean water tank capacity l 3,1
Waste water recovery tank capacity l 3.1
Water consumption per cycle
(minimum/maximum)
cm³ 420-990 420-1040 420-1190
Minimum water load cm³ 990 1040 1190
Maximum flow discharge rate (direct) cm³/s 30
Maximum discharge temperature (direct) °C 83
Absolute operating pressure (maximum) kPa 331
Electrical voltage and frequency V - Hz 230 – 50/60 (single-phase AC)
Maximum allowed voltage fluctuations % ±10
Absorbed maximum power W 2800
Heat emission J/h 0.3
Detected noise level in work areas under
suitable conditions (medium-peak)
dB(A) - dB(C) 56 - 63
Weight (mass) kg 62 63 68
Weight per support area with full tank and
maximum load
kg/cm² 1.51 1.53 1.63
Room temperature allowed °C from 15 to 40
Maximum relative humidity allowed % 85
Maximum altitude allowed with standard
safety valve
m asl 1500 (2000 with replacement of the valve)
Conditional connections:
maximum impedance
Ω 0.153+j00.096
REFERENCE STANDARDS
SterilClave sterilizers were conceived and designed in conformity with the following
directives and standards:
Directives
97/23/EC Pressure equipment
93/42/EEC Medical devices (class IIb) (and subsequent updates)
89/336/EEC Electromagnetic compatibility (and subsequent updates)
2006/95/EC Low voltage
Standards
EN 13060:2005 Small steam sterilizers
EN 61010-1:2001 Safety regulations for laboratory devices - Part 1: General regulations
EN 61010-2-040:2005 Safety regulations specific to sterilizers used in the processing of
medical material.
EN 61326-1:2007 Electromagnetic compatibility regulations for laboratory devices
Technical specifications