Clean the clean water tank 10 and the waste water recovery tank 11. Empty them
completely at least every six months and refill the clean water tank with clean water (see
Water supply table).
If the sterilizer will not be used for a long period, leave the tanks empty.
Filters
Replace the air sterilization filter 15 at least every one year: remove the filter and screw in
the new one.
Annually check the efficiency of the waste filter 20 as follows:
- open the door 16.
- remove the filter by pulling upward.
- check it, and clean with a jet of compressed air or replace it.
- replace the filter.
Checks
At least once a month check the efficiency of the door gasket 31 and replace it if there is
damage.
To extend the life of the filter, it is best to leave the door 16 ajar, but not locked with the
handle, when the unit is not being used.
Every day, check that there are no drips or seepage in the work area, since they can
cause a slipping danger.
Fuses
To calibrate the fuses 32, refer to the label plate inside the door 18.
At least annually, check the calibration of the pressure and temperature probes.
The probes check must be performed strictly by specialized personnel or at a
COMINOX TECHNICAL SERVICE CENTER
!