35
The SPECIAL cycle is an alternative to the standard cycles listed in the Table of
cycles/sterilizable material, and can be programmed based on specific sterilization needs
for special instruments or other loads. The SPECIAL cycle is not covered by the technical
requirements of the EN 13060 standard with which the COMINOX autoclaves comply: for
this reason, only qualified laboratory personnel are permitted to use the SPECIAL cycle.
These people, authorized by the Administrator, must execute a confirmation (for example,
in line with standard EN 554) for each special programmed cycle, with a specific
reference load.
3.2 Programmed start
Press ENTER to select the cycle to run. Use the ↑↓ arrows to select from among:
• 121 UNWRAPPED
• 134 UNWRAPPED
• 134 WRAPPED
• 121 HOLL./POROUS
• 134 HOLL./POROUS
• 134 PRION
• SPECIAL USER
• BOWIE & DICK TEST
• VACUUM TEST
• VACUUM+B&D TEST
The VACUUM+B&D TEST cycle calls for the Vacuum Test followed by the Bowie & Dick Test,
with a 5-minute pause between the two cycles and a preheating period to prepare the
sterilizer for the Bowie&Dick Test.
Press the
Å key to exit without programming any cycle to begin.
After you have selected the cycle, press ENTER to view the day. This value can be
changed using the ↑↓ arrows. Press ENTER to confirm the information and move on to the
next setting.
Once you have set the current date and time, check that the door 16 is closed and the
clean water tank 10 is full, and check whether the waste water recovery tank 11 needs to
be emptied. If any one of these operations has not been performed, the sterilizer will
indicate this with a message on the display.
Operation