EasyManua.ls Logo

ConMed ExcaliburPLUS PC - Page 10

ConMed ExcaliburPLUS PC
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Excel
use
of
a
medical
device.
Serious
injuries
ог
Ш-
ness
involving
the use
of
a
medical
device
must
be
reported
to
the
manufacturer
of
the
device
(or
to
the
FDA
if
the
manufacturer
of
the
device
is
not
known)
within
10
working
days
of
the
incident.
Summary
reports
of
such
injuries
must
also be
submitted
directly
to
the
FDA
twice
a
year.
Patient
deaths
relat-
ed
to
the use
of
a
medical
device
must
be
reported
to
the
manufacturer
and
the
FDA.
For
further
information,
please
contact
the
Regulatory
Affairs
Department
of
CONMED
Corporation
at
800-552-0138,
303-699-7600,
or
FAX
303-699-9854.
*
A
failure
of
the
unit
could
result
in
an
unin-
tended
increase
in
output
power.
*
Safe
and
effective
electrosurgery
is
depen-
dent
not only on
equipment
design,
but
also
on
factors
under
the
control
of
the
operator.
It
is
important
that
the
instructions
supplied
with
this
equipment
be
read,
understood,
and
followed
in
order
to
ensure
safety
and
effec-
tive
use
of
the
equipment.
*
Do
not
use
monopolar
electrosurgery
on
small
appendages,
as in
circumcision
or
finger
surgery,
as
it
can
cause
thrombosis
and other
unintended
injury
to
tissue
proximal
to
the
surgical
site.
Please
note
ANSI/AAMI
Standard
HF-18-1993
which
currently
con-
traindicates
both
monopolar
and
bipolar
elec-
trosurgery
for
circumcisions.
Urologic
literature
also
contraindicates
the
use
of
monopolar
electrosurgery
for
circumcision,
and
instead
recommends
the
use
of
bipolar
electrosurgery
where
clinically
indicated
for
hemostasis.
Physicians
choosing
to
use
bipo-
lar
electrosurgery
for
circumcisions
should
be
trained
in
this
technique
and
knowledgeable
of
the
effects
of
electrosurgery
on
tissue
of
this
nature.
Should
you
decide
that
the
bipolar
electrosur-
gical
technique
is
acceptable
for
circumcision,
do
not
apply
the
bipolar
electrosurgical
cur-
rent
directly
to
circumcision
clamps.
*
Apparent
low
power
output
or
failure
of
the
electrosurgical
equipment
to
provide
the
expected
effect
at
otherwise
normal
settings
may
indicate
faulty
application
of
the
return
electrode,
failure
of an
electrical
lead,
or
exces-
sive
accumulation
of
tissue
on
the
active
elec-
trode.
Do
not
increase
power
output
before
checking
for
obvious
defects
or
misapplica-
tion.
Check
for
effective
contact
of
the
return
electrode
to
the
patient
anytime
that
the
patient
is
moved
after
initial
application
of
the
return
electrode.
*
Studies
have
shown
that
smoke
generated
during
electrosurgical
procedures
may
be
harmful
to
surgical
personnel.
These
studies
recommend
using
a
surgical
mask
and
ade-
quate
ventilation
of
the
smoke
using
a
surgical
smoke
evacuator
or
other
means.
*
If
a
Dual
Pad
RETURN
Alarm
is
sounded
intraoperatively,
physically
confirm
proper
return
electrode
attachment
to
the
patient
prior
to
pressing
the
Monitor
Set
key.
Smooth
hand
over
pad
surface
to
ensure
electrode
con-
tact
to
patient
skin.
Replace
the
pad,
if
neces-
sary.
*
Simultaneous
activation
can
be
used
in
both
Standard
and
Spray
monopolar
coagulation
modes.
Caution
should
be
used
as
the
output
from
either
active
electrode
may
change
as
a
result
of:
activation
of
a
second
output,
end-
ing
activation
of
an
output,
or
changing
the
arc
length
from
an
output.
Power
sharing
is
unlikely
to
be
equal
because
of
differences
in
tip
to
tissue
distance
and
other
factors.
This
unequal
power
sharing
can
be
enough
to
stop
power
delivery
to
one
tip
if
the
second
tip
is
close
to
tissue,
and
the
first
tip
is
somewhat
above
the
tissue.
The
motion
or
deactivation
of
one
electrode
can
cause
the
other
electrode
to
start
delivering
power
when
it
had
been
too
far
away
from
tissue
to
arc
before
the
first
electrode
change.
It
is
recommended
that
a
second
electrosurgical
generator
be
used
as
a
better
alternative
when
it is
necessary
to
per-
form
simultaneous
operation.
*
Electrosurgical
leads
(active,
bipolar
or
return)
should
not
be
allowed
to
contact
the
patient,
staff,
or
other
electrical
leads
connect-
ed
to
the
patient.
©

Related product manuals