Exec
1.2
SPECIFICATIONS
INPUT
POWER:
700
W
max.
MI
rr
Mains
Voltage
Max
Mains
Current
Mains
Circuit
VRMS
ARMS
Breaker
Rating
Catalog
Freq.
Nom.
Min.
Max.
Nominal
Max.
Number
$
Hz
Mains
Mains
Freq.
(+1
Hz)
Idle
160-6290-120
60
120 104 127
0.8
7.0
10A/120V
60-6290-226
60
220
198
242
0.5 3.8
5A/240V
60-6290-230
50
220-230
198
253
0.5 3.8
5A/240V
|60-6290-240
50
240
216
264
0.4
3.5
5A/240V
60-6290-100
50-60
100
90
110
10
84
10A/120V
MAINS
OVERCURRENT
PROTECTION:
Combined
2-pole
circuit
breaker/power
switch
MAINS
FREQUENCY
LEAKAGE:
(Per
IEC
601-1)
Patient
Connections
to
Neutral:
Less
than
10
yA
Chassis
to
Neutral:
Less
than
25
wA
(120
V)
Less
than
45
A
(240
V)
TEC
Class
I,
Type
CE,
Defibrillator
Proof
4
v
VW
REGULATORY
CERTIFICATION
Non-Ionizing
Radiation
UL
Listed,
Medical
and
Dental
Eguipment
(E68077)
ISO
9002:1994
Registered
Facility
C
0123
POWER
DISPLAY
ACCURACY:
The
greater
of
+/-
10%
of
display
or
+/-3
watts
at
rated
load.
LINE
REGULATION:
Power
change
<
1%/V
for
VLINE
=
Min
to
Max
Voltage
Duty
Cycle:
15s on,
30s
idle
ENVIRONMENTAL
Operating
Conditions:
10°C
to
30°C,
95%
RH
Non-Condensing
maximum
MOUNTING
RESTRICTION:
2
inch
(5cm)
clearance
required
on
each
side
and
above
unit
for
cooling.
COOLING:
Natural
convection,
conduction
and
radiation.
No
fan.
Storage
Conditions:
-40°C
to
70°C,
altitude
-60
to
+4500
m
MSL
when
sealed
in
original
poly-
bag,
packing
material
and
shipping
carton.
Prior
to
shipment
or
storage,
the
unit
should
be
enclosed
and
sealed
in
a
polyethylene
bag and
placed
in
original
carton
using
original
packaging
materials.
a