ec
2.7.2
ACC
Faults
ACC
Codes
are
most
often
caused
by
faults
in
accessories
attached
to
the
unit.
Often
the
fault
can
be
corrected
in
the
operating
room
without
a
service
call.
The
following
list
gives
the
meaning
of
each
code,
and
what
to
check.
For
each
code,
check
to
see
that
the
switch,
panel
or
accessory
is
not
being
pressed
while
the
unit
is
powering
up.
Turn
unit
off,
unplug
accessory,
then
turn
unit
back
on.
If
the
code
still
shows,
call
service.
ACC
Code
What
to
check
FS
Monopolar
Foot
Switch
shorted
(Ensure
Foot
Switch
treadle
is
not
depressed.)
LH
Left
Hand
monopolar
accessory
shorted
tH
right
Hand
monopolar
accessory
shorted
bP
biPolar
accessory
shorted
*CPxx
Control
Panel
key
stuck
*Note:
“xx”
displayed
in
the
bipolar
window
cor-
responds
to
the
stuck
key
using
figure
2.1
as
a
reference.
Multiple
keys
show
as
“88”.
2.7.3 Err Faults
Err
Code
faults
generally
cannot
be
corrected
in
the
operating
room
and
require
a
repair
techni-
cian.
Occasionally,
turning
the
unit
off
and
back
on
will
clear
an
Err
fault
and
resume
normal
operation
of
the unit.
2.7.4
If
All
Else
Fails
Contact
CONMED
Corporation
Technical
Services
at
the
phone
or
fax
numbers
listed
on
the
inside
front
cover
of
this
manual.
Please
have
the
model
and
serial
numbers
from
the
rear
panel
nameplate
along
with
a
description
of
the
prob-
lem,
including
power
settings,
accessories
in
use,
and
fault
code.
2.8
ENVIRONMENTAL
PROTECTION
The
shipping
container
and
packing
material
should
be
retained
in
the
event
the
unit
must
be
returned
for
factory
service. At
the
end
of
the
equipment’s
life,
it
should
be
disposed
of
in
accordance
with
your
local
regulations.
Component
materials
are:
*
Aluminum
enclosure
and
heatsinks.
+
Unit
front
handle
is
steel.
*
Thermoset
printed
wiring
boards
containing
miscellaneous
electronic
components.
*
Power
transformer
made
of steel
and
copper.
*
Mains
cord
made
of
thermoplastic
and copper.
*
Shipping
container
is
cardboard;
packing
mate-
rials
are
a
combination
of
Urethane
foam
and
Polyethylene
film.

Related product manuals