ConMed ExcaliburPLUS PC - Unit Controls and Connectivity; Control Panel Details

ConMed ExcaliburPLUS PC
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.
Depress
the
Cut
treadle
of
the
monopolar
foot
switch.
The
Return
Alarm
tone
should
sound
and
Return
Fault
should
continue
to
glow.
While
holding
the
Cut
treadle,
rotate
the
volume
control
over
its
full
range
and
verify
there
is
no
signifi-
cant
change
in
sound
level.
Release
the
Cut
foot
switch
and
the
tone
should
stop.
7.
Verify
that
each
of
the
following
modes
can
be
selected
and
adjusted
up
and
down,
and
that
no
other
displays
are
affected:
*Pure
Cut
*Blend
1
Cut
*Blend
2
Cut
*Blend
3
Cut
*Standard
Coag
*Spray
Coag
*Bipolar
Coag
«Bipolar
Cut
Return
all
power
settings
to
desired
setting.
8.
Select
Bipolar
Coag
and
depress
the
bipolar
foot
switch.
Confirm
that
the
Bipolar
Mode
indi-
cator
illuminates,
and
a
coag
activation
tone
is
sounded, While
still
depressing
the
bipolar
foot
switch,
rotate
the
Sound
Volume
Control
over
it's
full
range
to
verify
the
function
of
that
control
and
that
sound
is
audible
at
all
positions.
This
control
may
be
left
at
any
desired
position.
Confirm
that
releasing
the
bipolar
foot
switch
returns
the
Excalibur
Plus
PC™
to
its
idle
state.
Verify
that
the
bipolar
hand
switch
control
is
functional
by
use
of
the
appropriate
forceps
or
by
connecting
a
jumper
between
the
center
and
bot-
tom
Bipolar
Output
jack
on
the
Output
Panel.
The
effect
will
be
identical
to
depressing
the
Bipolar
foot
switch.
9.
Press
the
Single
Pad
Select
key
to
select
Single
Pad
mode
and
confirm
that
no
ASSIST
Resistance
indicator
bars
are
lit,
and
the
Return
Fault
indicator
remains
lit.
Connect
a
single
pad
Return
electrode
to
the
Return
electrode
jack.
Confirm
the
Return
Fault
indicator
goes
out.
10.
Depress
the
Cut
treadle
of
the
Monopolar
foot
switch.
The
yellow
Cut
Mode
indicator
shall
light
and
the
cut
activation
tone
shall
sound.
11.
While
continuing
to
press
the
Cut
treadle
of
the
foot
switch,
depress
the
Coag
treadle.
An
operator
error
tone
shall
sound.
Release
the
Cut
treadle,
continue
to
press
the
Coag
treadle
and
confirm
that
the
tone
changes
to
the
coag
activa-
tion
tone,
and
the
blue
Coag
Mode
indicator
is
illuminated
until
the
Coag
treadle
is
also
released.
12.
Connect
a
handswitching
active
electrode
to
the
Monopolar
Accessory
Jack
#1.
Activate,
one
at
a
time,
the
Cut
and
Coag
hand
switch
controls
verifying
that
each
control
causes
the
correct
indi-
cator
and
tone
to
sound.
Move
the
handswitch-
ing
active
electrode
to
the
Monopolar
Accessory
Jack
#2
and
repeat.
13.
Remove
the
two
wire,
single
pad
return
elec-
trode,
and
confirm
that
the
Excalibur
Plus
PC™
beeps
twice,
and
the
Return
Fault
indicator
illu-
minates.
2.3.2
Preliminary
Performance
Testing
After
the
unit
passes
the
Preliminary
Functional
Tests
of
Paragraph
2.3.1,
preliminary
performance
testing
may
be
conducted.
Such
testing
is
best
carried out
by
use
of
an
electrosurgical
generator
tester,
as
described
in
this
manual.
Note
that
the
power
display
of
the
tester
will
be
most
accurate
when
a
noninductive
resistor
of
the
rated
genera-
tor
output
impedance
for
the
selected
mode
is
used.
If
no
tester
is
convenient,
the
availability
of
thera-
peutic
current
may
be
ascertained
subjectively
by
attempting
to
cut
and
coagulate
on
surrogate
tis-
sue
such
as
a
piece
of
meat
or
fresh
fruit,
a
wet
bar
of
soap,
or
a
sponge
moistened
in
saline.
2.4
CONTROLS,
DISPLAYS
AND
CONNECTORS
The
numbers
preceding
the
following
paragraphs
correspond
to
the
numbered
items
in
Figure
2.1.
2.4.1
Control
Panel
All
controls
except
for
the
sound
volume
control
and
the
power
circuit
breaker
are
located
on
the
control
panel.
Refer
to
the
CONTROL
PANEL
portion
of
Fig. 2.1.

Related product manuals