User Manual
29
Select the button【Default】, the print settings will back to the default.
In this interface, the automatic mode option can only be effective when select "auto" in 【Print
Mode】.
[Auto 3*4+1] /[Auto3*4]
/[Auto2*6+1] /[Auto2*6]/
[[Auto 4*3] /[Rhythm2]/
[Rhythm 3 ]/[Rhythm4]/
[Manual]
The selection will be used as the default print
mode.
The selection will be used as the default Lead
Gain. "Smart" means that the system will adjust
the lead gain automatic to fit the height of paper
while printing. "Current" means that the system
will use the screen lead gain while printing.
3Sec/4Sec/5Sec/6Sec/8S
ec/
10Sec/15Sec/20Sec/25Se
c
The selection will be used as the default print time
length of step.
10Sec/15Sec/20Sec/25Se
c/
30Sec
Under the print mode is "Rhythm 2", "Rhythm 3"
Or "Rhythm 4", the system will use the select
option as the print time length of rhythm strip.
[2*6]/[2*6+Mark]/[3*4]/
[3*4+Mark]/[4*3]/
[4*3+Mark]/[None]
Under the print mode is "Rhythm" Or "Auto",the
system will use the select option as the default
print style of average QRS.
The auto-diagnose contains 2 parts of data and
conclusion, user can print one of them only as his
wish.
[per1Min]/[per2Min]/
[per3Min]/[per5Min]/
[per10Min]/[per20Min]/
[per30Min]/[per60Min]/[off]
The system will print ECG periodically in the select
time interval, If the option of print mode is selected
as "Manual", the system will print in "Auto3 * 4 +
1". Otherwise, the system will print in selected
mode.
Fill the hospital name in it.
Note: Auto strip, Rhythm strip, Average QRS, Auto-Diag, Periodic are available when
"Auto" or "Rhythm" print mode is selected.
13.12 Checking Electrodes Placement
Select button on the main interface can start the following lead emplace illustration
interface: