64 10 ORIGINAL (ORG) MODE
4) Serial communication output format
Serial communication output format
5) Selection of Pressurization valve failure detection time・Differential pressure limit
Differential pressure limit
Pressurization valve failure detection
Verifies DPS offset during a pre-determined waiting time, upon completion of
the leak test, If DPS offset exceeds its limit, error message (E-9) is displayed
by regarding the status of AV1 (pressurization valve) as not closed.
Memory switch #23 needs to be set at “00” when the positive test pressure is
less than 10kPa or the negative one is less than –20kPa.