11 INTERFACE 77
4) Response
Command from host computer
Response of LS-1866
Buzzer
Read command
Data transmission
OFF
Write and control command
ACK(06H) transmission
OFF
Ineffective command or
command while tester is busy
Error transmission ON
Error response
#SS_00_CC_EE:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max.
Remarks
Identification No.
S
2 digit integer
00
99
Channel #
C
2 digit integer
00
15
Error field
E
2 digit integer
01
80
Checksum G 2 digit integer 00 FF
Data from “#” to “:” is summed up
and its twos complement is taken.
Error field
Code
Error
Description
01
Inappropriate data
Written value was inappropriate
10
Execution not available
Tester is not in a state to execute the command.
40
Checksum error
Checksum data does not agree.
80
Ineffective command
Command is inappropriate.
Transmitted data from LS-1866 by RLD command (Output format is same as T format.)
#SS_00_J_±VV.VV:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max.
Remarks
Identification No.
Symbol
2 digit integer
00
99
Judge field J
ASCII code
(1 digit
hexadecimal)
0 D
0: No test data
1:Lo NG
2:GOOD
4:Hi NG
9:LL NG
C:HH NG
D:ERROR
Data V
4 digit floating
point
Leak
rate unit
±0.000 ±0999.
Checksum G 2 digit integer
00 FF
Data from “#” to “:” is summed up
and its twos complement is taken.
Transmitted data from LS-1866 by other reading command
#SS_00_CC_±VVVV.VVV:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max.
Remarks
Identification No.
S
2 digit integer
00
99
Channel #
C
2 digit integer
00
15
Data V Fixed point
Leak
rate unit
±0000.000
±9999.000
Checksum G 2 digit integer
00 FF
Data from “#” to “:” is
summed up and its two
s
complement is taken.
11.5.2 Output formats
1) Operation
By the setting of memory SW, various settings can be made.
2) Baud rate setting
By the setting of memory SW_#18, the baud rate can be changed.
MSW 18
1
9600
0
19200
The channel # for the transmitted data for RTO is always 00because
RTO cannot be programmed independently for each channel.
NOTE