10 ORIGINAL (ORG) MODE 65
1
1
0
0
.
.
3
3
I
I
n
n
i
i
t
t
i
i
a
a
l
l
i
i
z
z
a
a
t
t
i
i
o
o
n
n
(
(
P
P
-
-
3
3
)
)
Initialization resets all settings to the manufacturers settings.
<Operation>
1) In the Original mode, press LEFT or RIGHT key several times until P-3 is displayed on the leak
rate indicator, and Init is displayed on the HI and LO leak limit indicators.
2) Pressing the ENTER key for 1 second or longer resets all the settings to the manufacturers settings.
10.3.1 Manufacturers settings
Memory switch
01:0, 23:00, etc:1
Limit
settings
BAL HI (WORK-side)
500(Pa)
gain coefficient
1.00
BAL LO (MASTER-side)
-500(Pa)
Leak Master setting
0.00 (mL/min)
DET HH (WORK-side)
55(Pa)
Noise reduction setting
OFF (0)
DET HI (WORK-side)
50(Pa)
Timers
Charge delay (DL1)
0.5(s)
DET LO (MASTER-side)
50(Pa)
Charge (CHG)
10.0(s)
DET LL (MASTER-side)
55(Pa)
Balance delay (DL2)
1.0(s)
Leak rate unit
Pa
Balance (BAL)
5.0(s)
Master-Preset compensation
OFF
Detection (DET)
10.0(s)
Equivalent internal volume (Ve)
0.000
End delay (DL3)
0.5(s)
Master-Preset value
0(Pa)
End (END)
0.5(s)
HH and LL ratio settings
10%
Stabilization (STB)
10.0(s)
1
1
0
0
.
.
4
4
C
C
h
h
a
a
n
n
n
n
e
e
l
l
C
C
o
o
p
p
y
y
(
(
P
P
-
-
4
4
)
)
Settings of a channel can be copied to another.
<Operation>
1) In the Original mode, press LEFT or RIGHT key several times until P-4 is displayed on the leak
rate indicator, and CoPy is displayed on the HI and LO leak limit indicators.
2) Press the ENTER key for 1 second or longer to copy the leak limit and the timer settings the current
channel to the next channel in sequence.
It takes about 1 second per copy. When copy to one channel completes, it goes on to the next
channel.
3) Pressing the STOP key copies to the current channel and cancel copying to the subsequent channels.
4) Press the STOP key again to return to Measurement mode.
This function (9.4) is only valid when the unit of leak rate is either Pa or kPa.
In case of using ml/min, please change it to Pa or kPa before carrying 9.4 out.
(Refer to 7.14 )