InstroTek
37
Where CR is count ratio (Average count/std. count)
determined by the gauge in the field, A and C are calibration
constants for the depth of measurement used in the field and
True density is density determined by a conventional method in
the lab.
Note: The gauge can perform the calculations if counts and
true density values are determined and input in the gauge
software. Furthermore, the gauge software is written so that the
entire special calibration function process, including collection
and storage of counts, is performed by the software. Follow
gauge instructions to perform Special Calibration.
To use the Special Calibration Function:
Press MENU:
Press 1 THEN 6
15. Buzzer Alarm
16. Special Cal.
UP/DOWN for Next
Select #, ESC Exit
Enable Special
Calibration?
Press YES or NO
ESC to Exit