Press YES to Enable Special Calibration and NO to go back to
the gauge ready screen.
Note: If special calibration has been used previously, the gauge
will ask if the same calibration or previously stored data should
be activated. At this point, the previously used special
calibration may be activated or select to start a brand new
special calibration.
At this point, the gauge provides the option to derive the Spec
Cal constant B or entering the constant previously derived.
When ‘Yes’ is selected, the gauge will prompt to select the
number of counts required to take on the material. It is
recommended that a minimum of four and maximum of ten
readings are selected for this test.
(Select 1 to 10 readings)
Use Gauge to Derive
B Value?
Press YES or NO
ESC to Exit
Enter Number
of Counts: ##
ENTER to Accept
ESC to Exit
Depth: BS
UP/DOWN TO CHANGE
YES to Accept
ESC to Exit