EasyManua.ls Logo

CTX NEXT - Operational Instructions; Instrument Overview; Main Menu Navigation

CTX NEXT
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Decommissioning
Operations to be performed every time the instrument is put out of service for long periods:
Switch off the instrument
Close the inlet and outlet valves outside the instrument
Check that there is water in the electrode holder so that the electrode is always wet. Otherwise loosen
the ring nut and put the electrodes in a container with some water or with some KCL solution.
OPERATIONAL INSTRUCTIONS
Instrument
The control station is equipped with an electronic control unit for the management of dosage; in Figure 13 is
shows the front panel of the control unit with firmware release.
Fig. 16 – Startup screen
Main Menu
After loading, the instrument display shows the following menu, depending on the version of the instrument in
your possession, the variables measured will appear on the display:
Fig. 3 Control panel Main Menu (ph RX Chlorine version)
Fig. 4a Probe inputs - Right side view of the instrument
1 Measurement panel
2
Menu key. Touch to access the
menu
3
No flow signal
3
Flow presence
signal
4
Alarm Signals
OVERoverdose
alarm
MAXmaximum
threshold alarm
MIN minimum
threshold alarm
5
Active Relay Signal.
1, 2, …. - indicates the number of
the active relay
ARROW indicates the type of
intervention, going up or down
1
2
3
4
5
ppm pH mV
58
ENGLISH

Table of Contents

Related product manuals