EasyManua.ls Logo

DATA PRECISION 1350 - AC Current; Functional Characteristics

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62-5008
MODEL
1350
Maximum
Current:
Series
2A
fuse
(250V)
Nominal
Full
Scale
Voltage
across
Shunts:
100mV
Accuracy
(1
year,
23°C
±5°C):
±0.5%
input
±1
l.s.d.
Temperature
Coefficient
(O-C
to
40°C):
±
.02%
input/°C
±
.01
%
range/°C
1.3.5
AC
CURRENT
Nominal
Range
Overrange
Resolution
100.0/liA
199.9iaA
O.VA
1.000mA
1.999mA
1
/xA
10.00mA
19.99mA
10/xA
100.0mA
199.9mA
lOOjuA
1000
mA
1999
mA
1mA
Maximum
Current:
Series
2A
fuse
(250V)
Nominal
Full
Scale
Voltage
across
Shunts:
lOOmV
RMS
Accuracy
(1
year,
23°C
±5°C):
30Hz
to
50Hz:
50Hz
to
500Hz:
500Hz
to
2kHz:
Temperature
Coefficient:
30Hz
to
2kHz:
±1.5%
input
±5
l.s.d.
±.75%
input
±2
l.s.d.
±1.5%
input
±5
l.s.d.
±.03%
input/°C
±.02%
range/°C
1.3.6
FUNCTIONAL
CHARACTERISTICS
Reading
Rate:
2.5
readings/second
Polarity:
Autopolarity
Overload
Protection:
Blanking
of
all
digits
leaving
sign
and
decimal
point
illuminated.
Overrange:
100%,
except
on
1000V
ranges
^Copyright
1977.
Data
Precision
Corporatton.
All
rights
reserved.
1-5