EasyManua.ls Logo

DATA PRECISION 1350 - AC;DC Converter Tests

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62-5008
MODEL
1350
START
XRia
CATHODE
WAVEFORM
R/R
01,
C16
R/R
Z7,C10,
014
TO
REFERENCE
SUPPLY
TROUBLESHOOTING
Z7-3
^
WAVEFORM
OK?
>
^^TP3
/^V
+
5%,
TP1
=
^.5V
+20%.
TP5
=
^\2.8V
+
20";^^
"H
R/R
T1,
CRU,
0R12
POWER
ON
Figure
5-4.
Power
Supply
Trouble
Shooting
Flow
Diagram
5.7
AC/DC
CONVERTER
TESTS
The
sequence
in
Figure
5-6
checks
the
operation
of
the
AC/DC
converter
of
the
Model
1350
DMM.
In brief,
a
standard
1
VRMS
signal
(a
500
Hz
is
connected
to
the
input,
and
the
performance
of
the
meter
is
checked
as
the
input
is
conditioned
to
a
proportional
dc
level
for
the
A/D
converter
input.
The
waveforms
to
be
used
as
standards
for
these
checks
are
included
in
the
waveforms
shown
in
Figure
5-8.
Note
that
the
first
check
is
a
dc
level
check
at
the
output
of
the
operational
amplifier,
Z1.
This
veri
fies
the
operability
of
this
active
component
by
observing
an
appropriate
average
dc
output
for
the
ac
input
applied.
©
Copyright
1977.
Data
Precision
Corporation.
All
rights
reserved.
5-7