AquaLab
7. Veri cation and Calibration
41
Correct
Not Correct
Next
Correct
Not
Correct
Measure Verification Standard
Correct
Not Correct
Repeat Process
Go to Linear
Offset Procedure
Go to Sampling
Procedure
Clean Sample
Chamber
Re-Read
Standard
Measure 2nd
Standard
Clean Sample
Chamber
is  owchart is a graphical representation of the directions given above
for checking for linear o set.