AquaLab
12. Troubleshooting
80
14. DUO PROBLEM:
Moisture Content displayed is not correct.
SOLUTION:
1) Model selected may not be correct for the product being test-
ed.
a. Toggle through the available models to  nd a more appropri-
ate model.
b. If the model is correct but not giving correct moisture con-
tent values it may be necessary to generate a new model for
the product or update an existing model. For information
about generating a model, contact Decagon Devices for up-
dating a model, refer to Chapter 10: Duo Operation.
15. DUO PROBLEM:
Moisture content does not show up on the screen.
SOLUTION:
Moisture content has not been activated.
1) Toggle to menu tab, select moisture content, and select the
appropriate model.
a. If no models appear in moisture content screen, models will
need to be reloaded using AquaLink RG software.
b. If moisture content is not an active selection, the DUO fea-
ture may not be active. Content Decagon Devices to learn
how to activate the DUO feature.