AquaLab
12. Troubleshooting
81
16. DUO PROBLEM:
Message on the screen displays the following:
SOLUTION:
1) When a moisture content reading is not shown, the water ac-
tivity or temperature for that reading is beyond the scope of
the moisture sorption isotherm.  is can happen under the
following conditions:
a.  e isotherm equation calculates a moisture content that
is less than 0% or greater than 100% with the given water
activity.
b.  e control temperature is signi cantly di erent than the
isotherm temperature.
Make sure that the samples water activity and the instrument’s
controlling temperature are within the scope of the selected mois-
ture sorption isotherm model.