CJC
The CPAD3-TH8-x comes with an integrated cold junction compensation sensor with an absolute accuracy
of ±0.2 °C. In order to archieve this accuracy the sensor has to be connected for at least 2 minutes to the
thermocouple connector (CJC equilibrium time).
NOTE: With the CPAD3-TH8-UNIVERSAL it is possible to get almost the same accuracy under laboratory
conditions compared to the CPAD3 with dedicated TC-connectors. If the environmental temperature is
rapidly changing, the accuracy may decrease three times more compared to the standard thermocouple
types! So the CPAD3-TH8-UNIVERSAL is not recommended for automotive measurements!
Programming information
The CPAD3-TH8-x programming information is available in the DEWE-MODULES Programmers Reference
Manual.