EPAD2 Module reset:
If the Module could not be detected from the software anymore a possible reason could be that the module
has been set to a different address or baud rate. With the reset function you can set the module back to its
default settings:
Baud rate: 9600 baud
Checksum: deactivated
Address: 0x00 (this equals a cleared module in Dewesoft)
Procedure: Press the ID button while powering on the module, and keep it pressed for at least 5 seconds.
CPAD2/CPAD3 Module reset:
If the module is not responding on the CAN bus anymore, or you have applied unwanted settings to the
module it can be easily set to the default settings.
Identier type: Extended
Data Identier: Bit 29 to Bit 25 = 0
Bit 24 to Bit 1: = Module Serial Number
Bit 0: = ChnBit (0 for Channel 0 to 3; 1 for Channel 4 to 7)
Data Format: Intel
Baud Rate: 500 kbaud
Sample Rate: 1 Hz
Procedure: Press the ID button while powering on the module, and keep it pressed for at least 5 seconds.
After that the module starts automatically measuring with 1Sample/Second. The status LED will indicate that
by blinking with 1 Hz. The module will send the data of the eight channels with two identiers that differ in the
last bit.
Default Data Identier Example:
Module Serial Number: 0341581
The module will send the data of channel 0 to 3 on identier:
341581*2= 683162 = 0xA6C9A
Data of channel 5 to 7 on identier:
341581*2+1= 683163 = 0xA6C9B
For detailed information how to change the identier refer to the Programmers manual.