EasyManua.ls Logo

Diatron Abacus junior 30 - Page 36

Diatron Abacus junior 30
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ROUTINE UTILIZATION & MEASUREMENT
33
MEASURE
New
Options / Lyse volume
Change lyse
These are the + or options you can
select from. See above description for
information.
Sampling depth
The analyzer requires a minimum of 2ml of whole blood in the sampling tube. AJ30
can however be adjusted for low volume samples. This becomes necessary when
there is extremely low volume of sample in the tube.
This option also allows using sampling tubes with an elevated bottom. In this case
you have to set a higher sampling level to avoid the needle hitting the bottom of the
tube.
MEASURE
New
Options / Sampling depth
Select the necessary option so that the
instrument can take sample from the right
location.
When all parameters are set, press the START button to save your settings and start
the measurement.
WARNING! Do not reach inside the instrument during operation,
as the needle can injure you!

Table of Contents

Related product manuals