M I N I - C U B E U S E R M A N U A L
Rev 1.3 06/2018 22/59
USE
Instructions for the operator
Before reading this section of the Manual, it is important to read the previous
chapters in order to avoid problems with the operation of the instrument.
Compatible test tubes
Standard 13 x 75mm, K2EDTA Hematology Tubes (BD)
K2EDTA Microtainer® Tubes (BD)
Sample requirements
The following sample requirements are critical for accurate results:
Minimum sample volume must be 3.0 mL in the standard 4.0 mL EDTA
Tube or 500 µL in the Microtainer Tube
The test must be performed within 4 hours of collection with the sample at
room temperature (18 25 °C).
The test may be performed on blood samples stored at 2 - C for a
maximum period of 24 hours. Ensure that the sample is at room
temperature for 15 minutes and well-mixed prior to inserting into the
instrument and performing the test.
Preparation of the sample
Before starting a test, the samples must be manually mixed as follows:
Gently and completely invert the test tube end over end 10 - 12 times
immediately before starting the analysis. Do not shake or agitate the sample
vigorously, as this could cause bubbling or haemolysis.
WARNING: Make sure the test tube is hermetically sealed.