M I N I - C U B E U S E R M A N U A L
39/59 Rev 1.3 06/2018
TROUBLESHOOTING
ERROR MESSAGES
CAUSE/REMEDY
Remove all test tubes
A cycle was started up with the test
tubes already inserted.
Remove the test tubes and start the
cycle again.
If the problem persists, there could be
a foreign body in front of the sensors.
Turn off the instrument, turn it upside
down and try to clean inside.
Err
A test tube has been removed from its
position before the end of the analysis,
or the system is having trouble
detecting the sample.
Low
The sample blood level too low (The
minimum volume is 2mL).
High
The sample blood level is too high or
there could be an issue with the
secondary label (number of labels or
thickness). Refer to Chapter 4 for the
ideal label placement.
Exam not available.
Need refill
No tests remaining. Refer to Chapter 4
to load a new transponder.
Printer error
Retry?
The printer could be OFF. Turn ON the
printer or repeat the installation printer
procedure.
USB pendrive not found
Remove and reinsert USB drive and try
again
CHECK DEVICE
already used
The transponder does not have any
tests remaining. Please load a new
transponder (Test Device Next).
Warning! Remain n tests
The number of tests remained is low
Tab. 6.0
Chapter
6