M I N I - C U B E – U S E R M A N U A L
Rev 1.3 06/2018 34/59
From the Settings menu, you can customize various functions:
BLUETOOTH: Enable or disable the Bluetooth connection with other devices.
BLUET. VISIBLE: Enables visibility for Bluetooth connections such as a
Smartphone.
AUTO PRINT: Set to “ON” for automatic printing of results at the conclusion
of the test.
ORGANIZE: Edit how data will be saved and displayed in the archive (by
name, surname, sample ID, patient ID).
DAS 28: Enable this setting to calculate the DAS 28 for analysed samples.
LANGUAGE: Select language, options are English, Italian, Dutch or Spanish.
TEMP. SCALE: Choose from Celsius or Fahrenheit.
DATE/TIME: Set the date and time format.
REFERENCE: Indicate the reference scale to be used for test results
(Westergren or Panchenkov).
TEMP. CORR.: Enable or disable Temperature Correction. See chapter 1 for
explanation of temperature correction.
THRESHOLD VALUE: Set a threshold value for the normal ESR reference
range limit; when this value is exceeded the instrument will display the result
in red to alert the operator that the sample exceeds the normal range (Default
is 20 mm/hr).
QC: View the QC menu and configure quality control settings.