UltiMate 3000 Nano and Cap - System Installation and Application
53
4.4 Synchronizing the Autosampler with a Pump
If your system includes a WPS-3000 or WPS-3000T autosampler, you can synchronize the
injection command of the autosampler with the strokes of a DGP-3600 or LPG-3400 pump.
Synchronization ensures that all injections are performed at the same phase of the pump
cycle.
On the Segments & Pump Link tab page of the autosampler's properties, determine which
synchronization is performed.
Please note: Dionex recommends always enabling synchronization as synchronization
helps enhancing the analytical results considerably.
Click the arrow in the Pump Link section and select an option from the drop-down list:
Fig. 28: Segments tab page
(here: when the UltiMate 3000 system includes an DGP-3600 pump)
Please note: The options are available only if both the autosampler and the pump are
installed in Chromeleon.