UltiMate 3000 Nano and Cap - System Installation and Application
67
9 Technical Information
Column flow range:
50 nl/min to 2.5 ml/min using predefined splitter
cartridges and active flow control
Flow control:
Electronically controlled and actively balanced flow
splitter, independent from solvent composition and
column backpressure
Gradient delay time:
Typically less than 1.5 min to splitter outlet at 200
nl/min
Sample thermostatting:
4 to 45°C, max. 22 °C below ambient
Thermostatted column
compartment and switching
valves:
5–70 °C (max. 15 °C below ambient)
UV detection:
Dedicated flow cells for nano-, capillary- and micro
HPLC with 10 mm path lengths
User input and display:
Large LCDs, programmable to show important system
parameters
Soft keys for operation during installation and
maintenance
System control:
All functions and parameters are software controlled via
USB 1.1 connection
GLP features:
In Chromeleon:
Automatic equipment qualification (AutoQ), System
Wellness monitoring and recording of all system
parameters in audit trail by Chromeleon software
MS software interfacing:
Via the appropriate add-ons to Chromeleon,
single-point control for
Analyst
®
(Applied Biosystems/ MDS Sciex)
HyStar
™
(Bruker Daltonics)
Xcalibur
®
(Thermo Electron Corporation)
Technical information: May 2006
All technical specifications are subject to change without notice.
For information about the technical specification on the individual system modules, refer to the
Operating Instructions for the instrument.