UltiMate 3000 Nano and Cap - System Installation and Application
13
3 Application Examples and Fluid Connections
Different system configurations allow you to use the UltiMate 3000 system for various
applications. For information about the related system setup and the fluid connections in the
system, refer to:
• Manual Injection (→ page 16)
• Direct Injection (→ page 20)
• Preconcentration (→ page 24)
• 2D LC Salt Plugs (→ page 28)
• Comprehensive 2D LC (→ page 33)
• Parallel LC (→ page 39)
When installing the fluid connections in an UltiMate 3000 system, keep the following in
mind:
Important: Different fitting systems are used in an UltiMate 3000 system. Therefore,
install the capillaries and fittings only at the positions for which they are
intended.
Important: Différents types de raccords sont utilisés dans le système UltiMate 3000.
Par conséquent, installez les capillaires et les raccords uniquement dans
aux endroits où ils sont prévus.
Important: When you operate biocompatible system modules, observe the instructions
for connecting the capillaries and other components. Refer to the related
sections of the Operating Instructions for the module(s).
Important: Avant de accorder les capillaires où des autres éléments à un module
biocompatible, référez-vous aux sections correspondantes dans le mode
d'emploi pour le(s) module(s).
Important: Dionex recommends starting the fluid connections for the system modules
at the pump. Connect the capillary to the pump first and rinse thoroughly
before connecting the other end of the capillary to another module.
Important: Lorsque vous connectez les capillaires, respectez la séquence suivante:
Raccordez le capillaire à la pompe. Rincez le capillaire avant de raccorder
l'autre fin du capillaire à l'autre unité.
Important: When connecting the capillaries, make sure that the connectors are free
from contaminants. Even minute particles may cause damage to the flow
splitter, flow control valve, and column.