UltiMate 3000 Nano and Cap- System Installation and Application
54
• Select None to disable synchronization.
• When the UltiMate 3000 system includes a DGP-3600 pump:
Select UM3PUMP _L_STRK to enable synchronization with the left pump. To enable
synchronization with the right pump, select UM3PUMP _R_STRK.
• When the UltiMate 3000 system includes an LPG-3400 pump:
Select UM3PUMP _STROKE to enable synchronization.
Please note: You can change the default synchronization assignment for a specific
application, using the PumpDevice and SyncWithPump properties in the
Commands dialog box under Sampler on the control panel or in the
program file (PGM).
To disable synchronization for a specific application, set SyncWithPump
to Off. Use the PumpDevice property to select the pump for which
synchronization shall be performed. (The drop-down list contains names
of the pumps installed in the Server Configuration program.)
Changing the settings for these properties from the Commands dialog box
does not change the standard synchronization setting on the Segments tab
page.
For more information, refer to the Operating Instructions for the autosampler and/or pump or
to the Chromeleon Online Help.
4.5 Displaying the Degasser-Related Parameters
If your system includes an SRD-3x00 solvent rack, the degasser can be controlled with
Chromeleon via the pump. In this case, the following degasser-related parameters are listed in
the Commands dialog box under the pump to which the solvent rack is connected:
Chromeleon Description
Degasser Turns the solvent rack degasser on or off.
For pumps operated with an SRD-3000 solvent rack, the setting
should always be On. For pumps that are operated without degasser
or with an external third-party degasser, select Off.
DegasserVacuum Indicates the status of the degasser vacuum in the solvent rack
(OK or NotOK; read-only).
SolventRackLeak Indicates the status of the leak sensor in the solvent rack
(Leak or NoLeak; read-only).
Please note: Dionex recommends always leaving the degasser on.
Turning off the pump automatically turns off the degasser, too.