CAUTION
⚫ Do not directly pour water, neutral detergent or disinfectant solution into the
rotor chamber. Otherwise fluids may leak into the drive units and cause
corrosion or deterioration to the bearings.
⚫ If the rotor chamber needs cleaning, clean with cloth or sponge moistened with
a neutral detergent solution. Sterilize the centrifuge by wiping with a cloth
moistened with 70% ethanol solution.
(3)Drive shaft
⚫ We recommend regular maintenance for drive shaft. You can wipe the drive
shaft with soft cloth, and then apply a thin coat of silicon grease.
(4)Door
⚫ Clean and sterilize the door using the same method as the step (1) above.
(5)Rotor
⚫ To prevent corrosion, remove the rotor from rotor chamber. If not in use for a
lone term, then detach the rotor lid and turn upside down to dry the tube holes
and keep clean.
⚫ For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of
silicon grease to the rotor when it is completely dry.