35
11.2 Tubes
11.2.1Cleaning and sterilizing tubes
Condition Material
PA
PC
PP
Cleaning
Cleaning fluids
AcidicpH5 or lower
X
X
X
Acidichigher thanpH5
O
O
O
Alkalinehigher than pH9
O
X
O
AlkalinepH9or lower
O
O
O
NeutralpH7
O
O
O
Warmwater(up to 70)
O
O
O
Ultrasonic cleaning
Neutral detergentpH7
O
O
O
Sterilization
Sterilization
Autoclaving
1150.7kg/cm230minutes
O
O
O
1211.0kg/cm220minutes
X
O
O
1261.4kg/cm215minutes
X
X
X
Boiling
15 to 30minutes
O
O
O
Ultraviolet sterilization
200-300nm
X
X
X
Gas sterilization
Ethylene oxide
O
X
O
Formaldehyde
O
O
O
PA: PolyallomerPCPolycarbonatePPPolypropylene
11.2.2Cleaning PC tubes

Related product manuals