29
1Turn on the POWER switch, the door lock will release automatically.
2The door lock will release automatically once the operation finished.
3It is available to release the door by pressing key once the rotor stops.
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available
to be opened manually.
1 Ensure if the rotor has stopped rotating.
Listen carefully to ensure no rotating sound can be heard.
2Insert a hexagonal spanner (same as the rotor spanner) into the unlocking
hole on the right side of the centrifugerotate 180° clockwisethe cover can be
opened.
10.3Replacement of fuses
1There are two fuses, 250V, 12.5A time-delay type, size: Ф5×20.
2The fuse holder is located in the power inlet. Pull out the fuse holder from
power inlet and replace the fuses if necessary.