22
8.2 RCF Operation
1Turn on the power switch.
2Seta RCF(Relative Centrifugal Force) value.
CAUTION
Do not exceed the allowable maximum RCF value of the rotor and adapters.
Press the key and choose speed unit ×g, the speed symbol will flash into
RCF value input status.
If no key is pressed after the speed value has flashed for 5 seconds, the input
mode will be shut down.
Rotate the key to input a RCF value, RCF increment is100×g.
3Set operating conditions
The other operation, please refer to the section 8.1.
8.3 Pulse Operation
This function is used to remove the residual samples adhered on the interior of the
tubes.
NoteThe keyworks only when the rotor stopped and the door is locked.
1Turn on the power switch and load the rotor to the shaft, tighten the rotor lid
and make sure it is in secured position, and then close the door.