thin coat of silicon grease on the rotor surface after drying.
4It is necessary for a regular maintenance for rotor, recommend to clean it each
3 months to keep cleaning of tube hole and shaft hole, and then apply a thin coat
of silicon grease on it.
11.1.4Autoclaving
All rotors are manufactured in high-strength aluminum alloy material or stainless
still and can be autoclaved: 121 (1.0kg/cm
2
), 20 minutes.
But some adapters are made of plastics, these adapters can be deformed after
autoclaving, so you'd better use other disinfecting methods.
11.1.5Bio-safe seal ring
The rotor is sealed by bio-safe structures, achieved using three high-temperature
rubber seal rings. The seal rings may fall off or aging after several autoclaving,
need to be replaced or re-installed. The replacement methods please refer to the
section 9.3.