Slide the cover plate B, noses first, over the red 2-pin connector H.
Secure the cover plate.
Fig. 13: Mounting the cover plate
Connect the battery terminal to the battery.
Mount the cover of the battery compartment.
Mount the power supply unit.
Switch the Evita on.
Enable the “DC module” option by entering “4811”.
Check the Evita by following the instructions in the Evita 4/Evita 2 dura Test Certificate.
IMPORTANT:
If the Evita should not start, the batteries might be exhausted. Connect the Evita to the power
supply and allow the batteries to recharge (several hours).
+