5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 15
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
• Slide the cover plate B, noses first, over the red 2-pin connector H.
• Secure the cover plate.
Fig. 13: Mounting the cover plate
• Connect the battery terminal to the battery.
• Mount the cover of the battery compartment.
• Mount the power supply unit.
• Switch the Evita on.
• Enable the “DC module” option by entering “4811”.
• Check the Evita by following the instructions in the Evita 4/Evita 2 dura Test Certificate.
IMPORTANT:
If the Evita should not start, the batteries might be exhausted. Connect the Evita to the power
supply and allow the batteries to recharge (several hours).
+
–
B
H