17 HPSV Controller PCB
17.1 Power Supply to the HPSV Controller PCB
X1
Pin Assignment Pin Assignment
5c +12 V 11c AGND
6c +15 V 12c DGND
7c –15 V
14c +5 V
CAUTION
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.
CAUTION
Risk of burns when operating the system with the rear panel (heat sink)
disassembled. Do not operate the system for longer periods of time with the