EasyManua.ls Logo

Dräger Evita 4 - 17 HPSV Controller PCB; Power Supply to the HPSV Controller PCB

Dräger Evita 4
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 86
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
17 HPSV Controller PCB
17.1 Power Supply to the HPSV Controller PCB
X1
Pin Assignment Pin Assignment
5c +12 V 11c AGND
6c +15 V 12c DGND
7c –15 V
14c +5 V
CAUTION
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.
CAUTION
Risk of burns when operating the system with the rear panel (heat sink)
disassembled. Do not operate the system for longer periods of time with the
rear panel disassembled.

Table of Contents

Other manuals for Dräger Evita 4

Related product manuals