5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 76
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
16 Pneumatics Controller PCB
16.1 Power Supply to the Pneumatics Controller PCB
16.2 Connector Pin Assignment of the Pneumatics Controller PCB
X1
Pin Assignment Pin Assignment
6c +15 V 11c AGND
7c –15 V 12c DGND
8c +24 V
14c +5 V
X2 (Background debug mode interface)
Pin Assignment Pin Assignment
1DGND5RESQ
2 BKPT 6 IFETCH
3 DGND 7 +5 V
4 FREEZE 8 IPIPE
CAUTION
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.
CAUTION
Risk of burns when operating the system with the rear panel (heat sink)
disassembled. Do not operate the system for longer periods of time with the
rear panel disassembled.