5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 34
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
5CO
2
Carrier PCB
5.1 Power Supply to the CO
2
Carrier PCB
X2 (Power supply only)
Pin Assignment Pin Assignment
1c +12 V 8c AGND
2c +24 V 15c DGND
3c +15 V 16c +5 V
4c –15 V
Power supply to the temperature hybrid and to the pressure sensor
Pin Assignment Pin Assignment
1 temperature hybrid +10 V-A 1 pressure sensor
+10 V-A
4 temperature hybrid AGND2 3 pressure sensor AGND2
CAUTION
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.