5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 23
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
2.5 Replacing the CPU 68332 PCB
Switch the Evita off.
Remove the CPU 68332 PCB.
Remove the battery from the old CPU 68332 PCB.
Remove the D22 module (EEPROM) from the old CPU 68332 PCB.
Remove the D22 module (EEPROM) from new CPU 68332 PCB.
Fig. 19: Component mounting diagram of the CPU PCB (replacing the CPU PCB)
Install the battery in the new CPU 68332 PCB.
Transfer the D22 module from the old CPU 68332 PCB to the new CPU 68332 PCB.
Check the jumper configuration (see old CPU 68332 PCB or, if necessary, the
functional description and the component mounting diagram).
Install the new CPU 68332 PCB in the Evita.
Secure the CPU PCB with the screws.
Load the relevant software into the CPU 68332 PCB.
Check the Evita by following the instructions in the Evita Test Certificate.
1
2
X3
1
X12
D22