5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 58
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
Remove the plug-in connectors from the Graphics Controller PCB.
Remove the Graphics Controller PCB.
Remove D29 and D40 from the old Graphics Controller PCB.
Mount D29 and D40 onto the new Graphics Controller PCB.
Mount the new Graphics Controller PCB by following the reverse method for
dismounting.
Load the corresponding software into the Graphics Controller PCB.
Check the Evita 4 by following the instructions in the Evita 4 Test Certificate.
10.4 Tests possible in DS mode
10.4.1 Diagnosis mode
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 3.2.1,
Diagnosis page “Microprocessor” of “Front”, page 64.
10.4.2 External DS mode using PC
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.5, Testing
“Front panel”, page 150.