NM2C DIAGNOSTICS (continued)
2.1.2 Service Mode
The following additional service mode screens are available by selecting the
appropriate Service Code:
PMS: For performing scheduled Periodic Manufacturers Service
INST: For initial installation of the NM2C. This selection establishes the start
date for the warranty period.
SRVC: For unscheduled service visits requiring component replacements
or calibrations.
The Service Mode screen appears as shown in Figure 2-3. Press the trigger to
highlight the first Technical Service Representative I.D. box; rotate the
selection wheel until the first character of the I.D. appears, then enter the
selection by pressing the trigger. Enter the remaining three I.D. characters in
the same manner.
The selection wheel can now be used to select the service screens described in
the following paragraphs. Selecting RESET and pressing the trigger will set
the last service date to the current date, and the hours run since last service
to zero. Selecting ORDM DISPLAY allows O.R. Data Manager diagnostic
information to appear on the screen (pressing the trigger while in the ORDM
mode returns the display to the Service Mode). Selecting EXIT and pressing
the trigger will return the display to normal operation.
MAIN SERVICE SCREEN
MACHINE SERIAL NUMBER : XXXXXXXX
LAST SERVICE DATE : 1- 3-94
HOURS RUN SINCE LAST SERVICE : 97
TOTAL HOURS RUN : 3383
SERVICE CODE SRVC
TECHNICAL SERVICE
REPRESENTATIVE I.D.
EXIT
NVRAM
DUMP
PMS
CRITERIA
MONITORS
SERVICE
LOG
MAIN
SELECTS
FUNCTION
TO ADJUST
SETTING
0 0 0 0
RESET
ORDM
DISPLAY
Figure 2-3: MAIN SERVICE SCREEN, SERVICE MODE