PMS PROCEDURE (continued) NM2C
6.14.4 Carbon Dioxide O.F.P.D. - if applicable
* 6.14.4.1 Set the gas selector switch to ALL GASES.
6.14.4.2 Open the cylinder of carbon dioxide.
6.14.4.3 Open one (1) oxygen cylinder valve.
6.14.4.4 Set the carbon dioxide flow to 500 ml; set the oxygen flow to 4
l/min.
6.14.4.5 Close the oxygen cylinder valve.
( ) 6.14.4.6 Does the flow of carbon dioxide cease when the oxygen pressure is
depleted? ___ (Y)
6.14.4.7 Close the carbon dioxide flow control valve.
* 6.14.4.8 Set the gas selector switch to O
2
+N
2
O.
6.14.5 (Deleted)
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. K
6-32
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS