Problem Solving
90
Instructions for Use Oxylog 3000 SW 1.n
!! Int. battery in use During ventilation, when the
external power source has
been disconnected the
internal battery becomes
the main power source.
When starting ventilation
while using the internal
battery this alarm will not be
issued.
Press the Alarm Reset key to
confirm the alarm.
11
!! Key failed Technical defect. To continue ventilation with this
device, verfiy the ventilation
settings and continuously monitor
the device functions.
Call DrägerService.
18
!!! Leakage
(not in NIV)
The measured expiratory
tidal volume VT is approxi-
mately 40% lower than the
inspiratory value.
Repair leaks in hose system and
possibly in the tube.
Use new flow measuring hoses.
9
Faulty flow sensor. Replace flow sensor.
The ventilator may not func-
tion properly.
Contact your local DrägerService
for additional support.
!! Loss of data Technical defect. Contact your local DrägerService
for additional support.
20
!! Loudspeaker
inop
Technical defect. To continue ventilation with this
device, verfiy the ventilation
settings and continuously monitor
the device functions.
Call DrägerService.
21
!!! MV high The upper alarm limit for the
minute volume MV has
been exceeded.
Check patient's condition, check
ventilation pattern, adjust alarm
limits if necessary.
8
Faulty flow sensor. Replace flow sensor.
The ventilator may not func-
tion properly.
Contact your local DrägerService
for additional support.
Alarm Cause Remedy Alarm
Rank

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 3000 and is the answer not in the manual?

Dräger Oxylog 3000 Specifications

General IconGeneral
Tidal Volume50 - 2000 mL
Ventilation ModesCPAP
Frequency5 - 80 bpm
Inspiratory Time0.2 - 5 s
PEEP0 - 20 cmH2O
Power Supply100 - 240 V AC, 50/60 Hz

Related product manuals