Instructions for Use Oxylog 3000 SW 1.n 89
Problem Solving
! Check settings
O2
The set O2 concentration
cannot be achieved with the
set flow.
Adjust inspiratory flow or O2 con-
centration (in accordance with
measured value).
27
!! Check settings
time
The expiration time result-
ing from the settings for
Freq. and I:E or Tinsp is not
possible.
Change Freq. or I:E or Tinsp.13
!!! Confirm device
OFF with rotary
knob
Key
O has been pressed
for 3 seconds.
To switch OFF: confirm.
To continue ventilation, press key
O again.
---
!!! Device failure Technical defect. Contact your local DrägerService
for additional support.
1
!! Flow measure-
ment inop
Measurement hoses for
flow measurement hoses
kinked, disconnected or
leaking.
Ensure flow measurement hoses
are connected correctly.
22
Flow sensor defective. Replace flow sensor.
Technical defect. Disconnect the patient from the
device and continue ventilation
without delay using another
ventilator.
Call DrägerService.
!! Gas delivery fail-
ure
Technical defect. Contact your local DrägerService
for additional support.
19
!! High frequency Patient breathes at a high
spontaneous rate.
Check patient's condition, check
ventilation pattern, correct alarm
limit fspn if necessary.
14
!! Int. battery
charging inop
Technical defect. Contact your local DrägerService
for additional support.
16
!!! Int. battery dis-
charged
The operating time for oper-
ation with the internal bat-
tery has expired and an
external DC supply has not
been connected.
The ventilator must immediately
be reconnected to a mains supply,
an on-board DC supply or a fully
charged battery.
2
Alarm Cause Remedy Alarm
Rank

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 3000 and is the answer not in the manual?

Dräger Oxylog 3000 Specifications

General IconGeneral
Tidal Volume50 - 2000 mL
Ventilation ModesCPAP
Frequency5 - 80 bpm
Inspiratory Time0.2 - 5 s
PEEP0 - 20 cmH2O
Power Supply100 - 240 V AC, 50/60 Hz

Related product manuals