Instructions for Use Oxylog 3000 SW 1.n 93
Problem Solving
Messages in the information window
(Numerical examples)
Message Cause Explanation/Remedy
f = 12 per min or
VT = 800 mL
I:E = 1 : 1.5
Flow = 15 L/min
Change in Tinsp, f or VT in ventila-
tion mode SIMV.
f = 12 per min or
VT = 800 mL
Tinsp= 0.7 s
Flow = 35 L/min
Change in I:E, f or VT in ventilation
mode IPPV (CMV), IPPVAssist
(CMVAssist).
I:E = 1 : 1.5
Texp= 2 s
Change in Tinsp or f in ventilation
mode BIPAP
®
(PCV+).
PEEP >10 mbar has been set but
not confirmed.
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Gas consumption = 10 L/min Standard display in information
window for the current gas
consumption.
(Battery capacity)
Standard display in information
window for the current battery
capacity.
Pinsp>=PEEP + 3 mbar Set PEEP + 3 mbar > Pinsp Set Pinsp > PEEP + 3 mbar
PASB = 22 mbar Change in ΔASB or PEEP. PASB
is the absolute pressure
resulting from PEEP + ΔASB.
Confirm PEEP above 10 mbar ?